Sensor for in-situ determination of
organic content in aqueous solutions.
No reagents needed, measures directly
in the sample. No filtration needed.
Method: UV absorption measurement with
turbidity compensation. Runs only with
transmitter CSM750. Measurement up to
max. 8 g/l TSS.
Temperature range: 2 up to 60 oC.
Ambient temperature: -10...50oC,
Process temperature: 2...40oC,
-
Successor
-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections Technical excellence Simplicity - F
- L
- E
- X
Fundamental selection
Meet your basic measurement needs
Technical excellence Simplicity - F
- L
- E
- X
Lean selection
Handle your core processes easily
Technical excellence Simplicity - F
- L
- E
- X
Extended selection
Optimize your processes with innovative technologies
Technical excellence Simplicity - F
- L
- E
- X
Xpert selection
Master your most challenging applications
Technical excellence Simplicity Variable
-
-
Characteristic
Sensor for SAC, dissoveld organic load
-
Size
l = 360 mm, Ø 127 mm
l= 14.04 inch, Ø 4.95 inch -
Design
Sensor head Stainless steel 1.4571 (AISI 316 L)
Sensor optical windows Quartz glass
Sensor housing POM -
Process temperature
5°C ... 60°C
41°F ... 140°F -
Ambient temperature
-20°C ... 60°C
-4°F ... 140°F -
Process pressure
pressureless < 0,2 bar
3psi -
Sample flow rate
in flow through armature min. 10 ml/min
-
Consistency of the sample
depending upon type of probe clear water to ca.1,0 g/l TS, cloudywater to approx.. 4 g/l TS
-
Specials
with an immersion assembly with pendulum frame directly installation in basins or with flowthrough armature behind sample preparation Stam°Clean
-
Application
Water / Waste water / Process water
-
Measuring range
0.3 ... 50 m-1 resp. 0.4 ... 60 mg/l COD traceable to KHP
15 ... 700 m-1 resp. 20 ... 900 mg/l COD traceable to KHP -
Consumables
no chemicals necessary